2026-07-07 sputter profilo
Process
Aluminium sputter, power = 115W, duration = 5min, 15min, 30min
Samples
Sample 1: 5min, 115W. 38-47nm

comments: very dirty - black spots on both Si and deposited Al
Reading 1:
Scan length: 240um Step height: 47nm Profile:

Reading 2:
Scan length: 80um Step height: 38nm
Profile:

Sample 2: 15min, 115W.
Reading 1:
comments: sample was much cleaner visually
scan location:

scan length: 325um scan duration: 325s horizontal resolution according to software: 3nm/pt

step visible at 0.12mm, but no flat region on either side
after fitting two point linear:
TODO
Last updated
Was this helpful?